Basics of High Voltage Probe Design

Contents:

[Document Version: 1.21] [Last Updated: 05/25/1998]


1. About the Author & Copyright

Basics of High Voltage Probe Design

Author: Samuel M. Goldwasser
Corrections/suggestions: | Email

Copyright (c) 1994, 1995, 1996, 1997, 1998
All Rights Reserved

Reproduction of this document in whole or in part is permitted if both of the following conditions are satisfied:

  1. This notice is included in its entirety at the beginning.
  2. There is no charge except to cover the costs of copying.





2. Personal Safety


Read the associated document: "Safety Guidelines for High Voltage and/or line
power equipment" before attempting to work with high voltage systems.  High
voltage can jump amazing distances when you least expect it.  The direct or
indirect consequences of this can ruin your entire day or a whole lot more.


3. Basic Considerations


Warning: DMMs may not be particularly forgiving of voltages on their inputs
exceeding their specifications.  Autoranging DMMs may be even more likely to
blowout as they are selecting the correct range - if there even is one.
Depending on your electrical and mechanical components, the chance of excess
voltage due to arcover, leakage, or component breakdown may be a major
consideration.  My analog VOM has survived many close encounters with HV.  You
should not assume the same for the typical low cost or even expensive DMM.
There is a reason for the high cost of commercial HV probes - these kinds of
factors are incorporated (hopefully) in their design.

A simple high voltage probe for a DMM or VOM may be constructed from a pair of
resistors.  This is suitable for DC measurements but without compensation,
will have a unknown AC response due to the very high impedance and stray
capacitance forming a filter - low pass or high pass depending on the amount
of stray capacitance and input capacitance of your meter or scope.  However,
this simple design is sufficient for the majority of consumer electronics work
which are mostly DC measurements.  I have not characterized the AC response of
this probe design.  However, if there is AC riding on your high voltage, it
may mess up your readings if there is no compensation provided as it may act
as a high pass filter.

To design the voltage divider, the input impedance of the meter must be taken
into account.  There is a minor but significant difference between DMMs and
VOMs.

DMM: Z-in is usually constant, often 10 M ohms.

VOM: Z-in is the voltage range (full scale) times the ohms/volt rating of the
meter.

Here is the basic circuit:

  High Voltage <------/\/\/\/\/\/--------+-------------> + to DMM/VOM
                           R1            |                      |
                                         \                      \
                                      R2 /                   R3 /
                                         \                      \
                                         /                      /
                                         |                      |
  Ground Clip  <-------------------------+-------------> - to DMM/VOM

R1 together with R2//R3 form a voltage divider where R3 is the internal
resistance of the DMM or VOM on the scale for which the probe is designed.

While R2 is not strictly needed, it is recommended that it be included and
approximately equal to the Z-in of the meter on the scale you will be using.
The reason to include R2 is to insure that high voltage never can reach the
meter.  The ground clip should be securely connected to the metal chassis of
the device being tested - the frame of a microwave oven or CRT
grounding/mounting strap of a TV or monitor - before it is powered up.  Both
R1 and R2 should be located in the probe head.

The only difficult part is locating a suitable resistor for R1 that has high
enough resistance and physically is long enough such that arcover is avoided.
One company that comes to mind has a name something like Victoreen - they
make/made radiation measuring instruments among other things that require
really high value resistors.  It can be constructed from several equal lower
value resistors in series if they are all approximately the same size.
Another possibility is salvaging the focus divider networks from dead flybacks
or TV/monitor voltage multiplier assemblies.  Even if the unit was discarded
as being faulty, where there are no internal shorts in the HV rectifier or
resistive network itself, the entire unit can be used intact.

In addition to basic safety precautions when working around high voltages,
some form of equipment protection should be considered in provide an arcover
path to ground should there be arcing over the surface of the resistor as well
as if the resistor should somehow decrease in value.  There is no telling what
can happen under less than ideal damp or dirty conditions.

A 'corona','arc', or 'discharge' ring could be placed around the resistor near
the low voltage end securely connected to the ground cable.  The idea is that
any arcing over the surface should find this as its destination before
obliterating your meter.

A variety of devices could be placed across R2 to limit the maximum voltage
present in the event of a breakdown.  Suitable devices include neon light
bulbs (NE2s without resistors); zener, avalanche, or ordinary diodes; or other
semiconductor junction devices.  Traditional surge suppressors like MOVs and
Tranzorbs may work but their off-state impedance may be too low compared to
R2).  The neon bulb is good since its impedance is essentially infinite until
its breakdown of 90 volts or so is reached.  In some cases, these devices will
be destroyed (semiconductors may short) but they will have served their
protective function and are a small price to pay to protect you and your
meter.


4. Frequency Response Considerations


Probe compensation similar to that used on oscilloscope probes can be
implemented.  However, the determination of the capacitor values is beyond the
scope of this note.  To put it simply, the ratio of the capacitance C1:C2
(where C1 is across R1 and C2 is across R2//R3) needs to be equal to the ratio
of R2//R3 to R1 (or equivalently, to the inverse of the voltage divider
ratio).  C2 includes the stray capacitance and input capacitance of the meter
or scope probe.  The capacitor across R1 would need to sustain the HV so that
is another complication.  Since a 10x scope probe usually has an input
impedance of 10 M, the same design as used for the DMM would work with a
scope.  Although I have not pursued this issue, it sounds like based on the
ratio (1000:1 would mean that C1 would need to be extremely small, probably
smaller than the stray capacitance of the R1 and the associated wire) you
would need to add a capacitance for C2 and that there will be enough stray
capacitance such that no physical C1 will be needed.

If you are only interested in DC measurements, putting a .1 uF capacitor
across R2 should smooth out any 50/60 Hz or higher frequency ripple.

The implementation of full probe compensation is left as an exercise for the
motivated student.

Design example: 50,000 V maximum using a 10 M ohm Z-in DMM.
----------------------------------------------------------

By my rule above, I will select R2 to be 10 M ohms.  Fine adjustment of
calibration could be made by making R2 out of a combination of a fixed
resistor and a multiturn pot.

To minimally load the circuit under test, R1 should be as high as practical.
Practical here means (1) low enough so that leakage over its surface is not a
problem, (2) low enough that a reasonable voltage can be developed across
R2//R3, and high enough so that loading of equipment being tested will not
change the readings by more than a few percent.

R2//R3 is 5 M ohm.  Selecting R1 to be 5,000 M ohm will give a 1000:1 ratio so
that 50,000 volts will read out as 50 V on the DMM.  5,000 M ohm is high
enough that loading of a 250 M ohm focus network should not be an issue (5%).
1000:1 is a nice easy to remember ratio.  You could go to something higher if
loading is still a concern but then leakage current over the surface of R1
becomes an even greater concern.  Even 5,000 M ohm is about as close to an
open circuit as you can get - any contamination whatsoever will change the
calibration significantly.  You may find that using a 1,000 M ohm resistor
will result in less of a problem and accept the circuit loading that this
value implies.

Design example: 50,000 V maximum using a 30K/V VOM.
---------------------------------------------------

This is a little more complicated because you need to pick a range and then
calculate the Z-in for that range.  for example, for the 100 V range of a
30K/V VOM, the Z-in will be 3 M ohms.  For the same 5,000 M ohm R1 and 10 M
ohm R2, you would get a reading of 23 V (roughly) on the 100 V scale for a 50
KV input.  The divide ratio in this case is about 218.

It is a simple matter to determine a scale and an R2 such that the actual high
voltage measurement is easily calculated from the meter reading.  What you
want is the ratio of R1 to R2//R3 to be a nice round number.  Note that
switching ranges will produce some peculiar behavior due to this current
division between R2 and R3.  A unique R2 must be selected for each range of
interest.  You are already using nearly the maximum sensitivity of the meter
and switching to a lower range will only slightly change the position of the
needle unless you construct a range switch box as shown below.


5. Calibration


Unless you have a calibrated HV supply, a working TV for which you have the
service manual makes a good starting point.  The proper high voltage is
usually specified to within 5-10%.  If you have a line-transformer based HV
supply (e.g., neon sign transformer, rectifier, capacitor), then this would be
pretty accurate based on your power line voltage.  For a DMM with a constant
input resistance, you can use a low voltage (like a few hundred V) on a lower
range and extrapolate for the HV range.  However, for a VOM, you cannot use
this technique since changing ranges also changes the parallel resistance of
R2//R3.  You are already using nearly the full sensitivity of the meter.


6. Sample Circuit


I have constructed a high voltage probe from the surplus bleeder resistor from
a defunct video terminal.  For the probe tip, I used a discarded probe from a
VOM.  The resistor and probe tip were mounted inside an insulating plastic
tube with R2 included at its base.  A ground cable with an allegator clip
provided the connection to the chassis.  A second pair of wires with banana
plugs connected to the meter via a switchbox which could select between a DMM
or a couple of different scales on a VOM.  Potting the entire HV head is a
good idea to minimize the possibility of arcover.  Remember that 50,000 V can
jump several inches (2 inches in dry air approximately).  See above text for
other suggestions on equipment/you protection (which is not shown).

This circuit uses only a 203 M ohm high voltage resistor.  Since the internal
resistance of a typical focus divider network is 200-300 M ohm, this probe
would obviously load such a circuit excessively.

                                       .
 High Voltage <----/\/\/\/\/-------+--------+-------------+-----o + to DMM/VOM
                      R1           |    .   |             |
                     203 M         |    .   \             |
               (15 W HV rated)     |    .   / R4          |
                                   \    .   \ 360K    SW1 |
                                R2 /    .   /            /
                                1M \    .   |         3 o o o 1
                                   /    .   |           | 2 |
                                   |    .   \           \   \
                                   |    .   / R5    R6  /   / R7
                                   |    .   \ 25K   1M  \   \ 810K
                                   |    .   / Adj       /   /
                                   |    .   |           |   |
 Ground Clip  <--------------------+--------+-----------+---+---o - to DMM/VOM
                                        .
                      Probe Head        .    Range Switch Box
                                        .

SW1 - SPDT Center Off Toggle Switch.
All resistors except R1 are 1/4 watt 2%.

50,000 V maximum based on the particular R1 I had laying around. 1000:1
voltage division.
                                              
SW1-1 = Radio Shack DMM on 3, 30, or 300 V scale (10 M Zin).
SW1-2 = VOM, 25 V scale (30K ohms/V, 750 K Zin).
SW1-3 = VOM, 100 V scale (30K ohms/V, 3 M Zin).


7. General construction comments


(From: Duane C. Johnson (redrok@pclink.com)).

1. When you make the resister strings make sure that the maximum voltage
   rating of the resisters is not exceded. Most resisters have a maximum
   rating of 200 volts. In this case I would not excede 100 volts per
   resister.

2. Don't place the scope directly in series with the higher resisters. The
   first resister should be in parallel with the scope and and ground or the
   scope chasis. 

   This detail is important for safety reasons. If the connection to the scope
   becomes disconnected then their will not be a dangerious shock hazard as
   would be the case if the scope was in series.

(From: Larry G. Nelson Sr. (nr@ma.ultranet.com)).

You can do this with a high voltage resistor divider network. That is what is 
in a high voltage probe you would buy. This can be very dangerous to you and 
your equipment in the event of a failure. Please be very careful. I suggest a 
fuse at the probe input and an MOV across the resistor to ground that will 
connect to the scope and use a plexiglass tube to put it all in to contain the 
bits if anything blows up.

(From: Kevin Astir (kferguson@aquilagroup.com)).

With respect to preventing high voltage arcing and corona, *do not* use RTV.

Places that carry the GC line will have some 'anti corona discharge dope'
often called 'Q-dope'.  This is *the* stuff to use at HV.  You can clean it
off with acetone when you discover that you didn't clean flux off good enough
and have an arc underneath.  Epoxy and RTV have no such advantage, and RTV
releases corrosive acid while curing to boot.

Heed the warnings of other respondants WRT resistor voltage.  As they said 100
V per for garden variety resistors will yield a safe margin.  200V is typical
max rating.

There are special HV resistors (up to 10 KV or so) made, available into the G
Ohm range.  I don't know of a hobbiest source however.  If you know anyone who
works in nuclear instrumentation field they may be able to snag one for
you. (HV used as detector and PMT bias in radiation detectors).  This is what
will be inside "real" HV probe from Fluke, or Tektronix.

Finally, I have a lot of experience, and am fairly blase around HV, but in
addition to "normal" 115V AC rules, (no rings, one hand in pocket, etc.) I
*never* work on HV stuff (not even a TV or hi-pot test) alone.  And, I make
sure the 'observer' knows CPR, even if I have to wait 2 days to fix TV, so
girlfriend can 'help'.


8. High voltage probe frequency response


(From: Winfield Hill (hill@rowland.org)).

You can calculate this for yourself.  The parasitic lead-to-lead capacitance
of a typical small resistor is 0.05 to 0.2pF.  The capacitance from the
*middle* of the resistor and from any connection node between series resistors,
to ground, may range from 1pF to 5pF or more depending upon your choice of a
sheilding scheme.  Longer glass resistors intended for high voltages have
lower lead-to-lead capacitance, but higher distributed parallel capacitance.

As a worst case, imagine a 1000 M ohm probe made with a 2-inch long resistor.
To start, place the capacitance to ground from the midpoint.  If you assume
5pF of parallel capacitance, you'll see you're in trouble even at 60Hz!

* HV probes with usable high frequency response may have cleverly arranged 
  sheilds which can be connected to ground, the attenuated signal output, 
  and the input.  Some shields may overlap and may be adjustable, to provide
  an adjustable capacitive divider that can swamp the remaining effects of the
  resistor's capacitance.

* These probes can be very accurate at DC and high frequencies, but the skill
  of the designer is tested in the transition between resistive and capacitive
  divider action.  They can be very badly behaved in this transistion region.
  Basically the "center" of the resistor experiences a bypass to ground (a
  severe dip in gain) or due to other HF components, a severe peak in gain.
  This can easily be 50% and cover a decade.  Typically it'll happen at 20 to
  100 KHz.

  One solution invokes the capacitance from a few carefully-placed concentric
  sleeves connected to the input and the signal output, plus an overall shield
  or guard connected to ground.

* Poorly designed probes can suffer from mid- or high-frequency errors and
  from severe pickup sensitivity to nearby rapidly-changing voltages.  For
  example, I've seen probes that provide an output even when the measuring tip
  is grounded!

The classic low-voltage probe architecture of a pair pf RCs doesn't work for
HV scope probes, unless (1) you're willing to have an overly high capacitive
loading, or (2) you don't care about mid-frequency or pulse-shape response
accuracy.  This is  because the RES1 value will be very high, 100M or more
likely 1000 M ohms, and physically long and large.  So the real circuit is
like:

                +------------ CAP1 -------------+
                |                               |
Probe tip o--------- Rs -*- Rs -*- Rs -*- Rs -------- etc
                         |      |      |
                         Cs     Cs     Cs
                         |      |      |
                        Gnd    Gnd    Gnd

Because the Rs are so high, the probe becomes a good antenna, and a shield is
mandantory.  Therefore the Cs "stray" capacitance is higher than you might
think.  I think you see the problem.  

One solution is to make C1 very large, but it's just a matter of specs - if
you want 1% performance over the whole range, C1 is a severe load.  There is
a good overall solution, which I think is fairly clever (after thinking of it,
I discovered the experts had beat me to it!).


9. High voltage probes for AC measurements


(From Winfield Hill (hill@rowland.org)).

The usual method is simply to use a capacitive divider, a small 1 kV
capacitor, etc., or make the HV capacitor yourself for really high voltages,
like 5 to 20 kV, use an air neutralizing capacitor, etc.

Say for example, its a 3pF capacitor.  With shields.  With another more
conventional capacitor, say 3000pF for the bottom of the attenuator, followed
with a voltage buffer if desired, and you've got a nice wideband 1000:1 HV
probe installed in the system, good for mucho kV.


10. Commercial high voltage probes


(From: Frank Miles (fpm@u.washington.edu)).

Both Tektronix and Hewlett Packard sell HV probes rated at 5kV and up.
Bandwidths are (at least) into the 100kHz area, probably more.  I imagine
there are others.

The older Tek probes even had ports to refill with now-banned chemicals.  The
newer ones don't, but are more expensive.

Frequency response is a significant concern.  Designing and manufacturing a
decent HV probe is definitely non-trivial if you need flat frequency response.
Many parts have significant voltage coefficients, too, as well as breakdown
voltages.

(From Winfield Hill (hill@rowland.org)).

A significant part of the design effort (and cost) deals with, the problem of
how to go smoothly from a resistive divider at low frequencies, to a
capacitive divider at high frequencies, while keeping a constant attenuation
value at mid-frequencies.  This isn't easy.  Consider for example, that an
overall shield is clearly needed and must properly prevent the high-Z end of
the probe from simply acting as an antenna (as some HV probes do! i.e.  ground
the tip of the probe and *still* see large signals at the output).  This
shield acts as a capacitance to ground for the HV resistor, routing some of
the high-frequency current which is supposed to go to the output, to ground.
Hence at some middle frequency there's a dip!  This is solved in various ways
- with shields connected to the probe tip (but inside the ground), capacitors
bypassing the resistor, special resistor construction, etc.  Most solutions
can just as easily cause a region with a response hump, as well as a dip, or
even both.  BTW, these problems are much harder if one seeks to make a probe
with very low capacitive loading and high frequency response.  The Tek P6015A
probe is 3pF, and you'll also note it has a veritable raft of response
adjustments on the scope-input end.

Much of the cost of the probe is knowing how to do all this!

Incidentally, a low-cost intermediate-range HV probe is the Fluke PM9100,
which is a 4kV 100:1 probe with a 200MHz bandwidth.  Also the Tek P5100 is
rated to 2.5kV.  Most of these probes also have a derating above some d
frequency.

Most of this mess you can avoid entirely by not attempting to make the probe
measure DC (or at least not the whole frequency range).


Written by Samuel M. Goldwasser. | [mailto]. The most recent version is available on the WWW server http://www.repairfaq.org/ [Copyright] [Disclaimer]